EKO INSTRUMENTS CO., LTD. I-V Curve Checker MP-11 Instruction Manual Ver.7
Controlling Measurements on PC using
MP-11 Main Unit, External Pyranometer, and Thermocouple
I-V Characteristics
Irradiance
Temperatures
Managing the measured data on PC
Using the measured data on another application
Centralizing the measurement control, data analysis and management on PC
Using high precision pyranometer installed at the site
A. MP-11 Main Unit
B. PV Probe
(1.5m, Red 1pc, Black 1pc)
C. Grounding Cable (3m)
D. Battery Receptacle Box or
AC Adapter (1.8m)
E. Sensor Unit
F. Moduler Cable (3m)
G. Thermocouple (3m, 2pcs
H. External Pyranometer*
I. Output Cable for the
External Pyranometer*
J. USB Cable (12)
K. PC Installed with MP-11
Software
*Not included in the MP-11
package
Disconnect the PV module/array, which is subjected to take measurement, from the
relating devices such as power conditioner completely.
Set the switch on the Sensor Unit rear panel to [EXT] position.
Connect thermocouple cables to the Sensor Unit and setup the other end of
thermocouple cable at the measuring position.
Connect an external pyranometer to the Sensor Unit.
Connect the Sensor Unit to the Main Unit.
Ground the MP-11 Main Unit with grounding cable.
Connect MP-11 Main Unit to PC.
Turn ON the Sensor Unit power.
Turn ON the MP-11 Main Unit power.
Turn ON the PC and startup the MP-11 I-V Checker Software
Setup the communication port between the MP-11 Main Unit and the PC
Synchronize the date and time between the MP-11 Main Unit and the PC