EKO INSTRUMENTS CO., LTD. I-V Curve Checker MP-11 Instruction Manual Ver.7
Table 10-2. MP-11 Software Error and Warning Messages - Continued
Definitions and Corresponding
Either the PV module/array is over range or PV probes are
connected in reverser polarity when [Start] button is clicked.
The Open Circuit Voltage and/or the Short Circuit Current for
PV module/array terminal to be measured are more than
1030V and/or 36A, which are over the MP-11 measureable
range. Stop measurement immediately; it may cause
damage to MP-11.
The PV probes are connected with reverse polarity
(plus/minus terminals), and the measurement is terminated.
Check the PV probe polarity and reconnect as necessary.
When [Set Clock] button is clicked, this message appears.
Communication error between the MP-11 and PC software, or
MP-11 failure had occurred.
Check the cable connection condition between the PC and MP-11
Main Unit; make sure the MP-11 power is turned ON then retry the
[Set Clock] button.
If this error message still appears, contact EKO for further
assistance.
This message appears when preparing report and the Format.xls
file cannot be found.
Make sure the Format.xls file is in the same folder as MP11.exe file.
If these 2 files are not in the same folder, this “Format file read
error” occurs. If this file is deleted, uninstall the I-V checker
control software and reinstall. (For installation and uninstallation of
the software, see [9-1. Software Installation and Uninstallation])