2.3 APPLY PRESSURE AND RECORD RESULT
1. Increase pressure by turning handle clockwise, slowly and smoothly, until either of the following end
points:
•For destructive testing, the dolly and coating pulls off the substrate.
•For non-destructive testing, the minimum specified pressure value is reached.
If possible, complete the test within 90 seconds of starting. This is in accordance with some adhesion
testing standards.
2. Record the following information:
Pressure - indicated by the gauge
Test location
Type of adhesive
Substrate/adhesive temperature
Air temperature
Relative humidity
Dewpoint temperature
Coating system details
Duration of tests
Appearance of breaks, e.g. clean between coating and substrate, separation of coating layers,
jagged edges