12
13) "GSt.n" - significance of the zone of return on settings operation.
14) "EnU.n" - sanction of settings operation.
"EnU.n" = 0 - the settings operation on the channel n is forbidden;
"EnU.n" = 1 - the settings operation on the channel n is authorized.
Factory installation(aim) "EnU.n" = 1.
15) "rL1.n" - condition of the setting relay I at breakage of input source circuits of the
channel n.
"rL1.n" = 0 - is switched off;
"rL1.n" = 1 - is included.
Factory installation(aim) "rL1.n" = 0.
16) "rL2.n" - condition of the setting relay II at breakage of input source circuits of the
channel n.
"rL2.n" = 0 - is switched off;
"rL2.n" = 1 - is included.
Factory installation(aim) "rL2.n" = 0.
17) "rL3.n" - condition of the setting relay III at breakage of source circuits of the channel n.
"rL3.n" = 0 - is switched off;
"rL3.n" = 1 - is included.
Factory installation(aim) "rL3.n" = 0.
18) "rL4.n" - condition of the setting relay IV at breakage of source circuits of the channel n.
"rL4.n" = 0 - is switched off;
"rL4.n" = 1 - is included.
Factory installation "rL4.n" = 0.
19) "PP1.n" - minimum significance of a range of record of measured quantity.
Factory installation "PP1.n" = " 0.0".
20) "PP2.n" - maximum significance of a range of record of measured quantity.
Factory installation "PP1.n" = " 100.0".
21) "CLr.n" - line resistance calibration at measurement of the THERMOCONVERTER OF
RESISTANCE on 2-circuits diagram, expressed in Оhm:
- Connect a line closed on the part of the THERMOCONVERTER OF RESISTANCE to the
device or connect a resistance, equivalent to the line resistance;
- choose the appropriate channel in article "CLr.n" of the menu and press the button ">"
(the installed line resistance significance will be highlighted on the indication;
- press the button ">" (indication "no");
- press the button "^" (indication "YES");
- press the button ">" (the measured significance of the line resistance will be highlighted in
some seconds);
- press the button "^" for exit in the basic menu or «PGM» for exit from the menu.
22) "CLc.n" - calibration resistance of the cold seal compensator at measurement of the
THERMOCONVERTER OF RESISTANCE, expressed in ohm, in 0°С:
- connect the thermocouple and the cold seal compensator to the device, according to fig.
7.2 and put the working end of the thermocouple into an ice-water mix (0°С);
- choose the appropriate channel in article of the menu "CLc.n" and press the button ">"
(the installed compensator significance R0 will be highlighted on the indication);
- press the button ">" (indication "no");
- press the button "^" (indication "YES");
- press the button ">" (the measured value of the compensator resistance in 0°С will be
highlighted in some seconds); (in some seconds the measured significance of the equalizer will be
highlighted;