Elma GmbH & Co KG Kolpingstr. 1-7 D-78224 Singen Tel. +49(0)7731/882-0 Fax +49(0)7731-266 info@elma-ultrasonic.com
Service Manual
Transsonic Analog
Transsonic LC
Ultrasonic Cleaning Units
y english y
www.elma-ultrasonic.com

Table of Contents

Questions and Answers

  • A
    Andrew DoyleAug 8, 2025
    Why does Elma Other have no ultrasound but heating is OK?
    • C
      Christine GarciaAug 8, 2025
      If there's no ultrasound but the heating is working, the issue could be with the interference filter coil or the timer. In such cases, consider replacing the component or the complete interference filter.
  • J
    Jacob TateAug 12, 2025
    What to do if Elma Other has no ultrasound and no heating?
    • K
      Kristin HernandezAug 12, 2025
      If there is no ultrasound and no heating, the issue might be a faulty printed board (BZW diode; transistors; diodes mains and HF). You can try replacing the component or the complete printed board. Alternatively, disconnect the heating cable and check the unit again with new fuses. If the short circuit doesn't occur again, the heating is faulty, and you should replace it.
  • G
    gonzaleztroyAug 17, 2025
    What to do if Elma T700 Other fuse is released, with no heating and no ultrasound?
    • E
      Eric MonroeAug 17, 2025
      If the fuse is released, and there's no heating or ultrasound, it indicates a short circuit in the heating system. Disconnect the heating cable and check the unit again using new fuses. If the short circuit does not occur again, it confirms that the heating is faulty. Replace the faulty heating.
  • G
    glopezAug 23, 2025
    Why does Elma T700 Other have no ultrasound or weak ultrasound, but heating is OK?
    • V
      Virginia FloresAug 23, 2025
      If there is no ultrasound or weak ultrasound, but the heating is okay, the issue could be with the output transformer, the output capacitor (6.8nF), or scorching in the transducer system. You may need to replace the component or the complete printed board, or open and check the transducer system.
  • S
    Scott UnderwoodAug 28, 2025
    What to do if Elma T700 Other has no ultrasound but heating is OK?
    • M
      Meredith ArnoldAug 28, 2025
      If there's no ultrasound but the heating is working, possible causes include a faulty printed board (BZW diode; transistors; diodes) or a short circuit in the heating. Disconnect the heating cable and check again. You might need to replace a component, the complete printed board, or the heating element.
  • K
    Karen BakerSep 1, 2025
    What to do if Elma T700 has no heating but ultrasound is OK?
    • C
      Craig TorresSep 1, 2025
      If there is no heating but the ultrasound is working, the issue could be a faulty heating switch. If that's the case, replace the component.

Summary

General Information and Safety

Important Safety Warnings

Crucial safety precautions to be observed during operation and maintenance.

Important Operating Instructions

Intended Use and Safety Checks

Specifies the intended application and procedure to inspect the unit for damage.

Description of Operating Elements

Technical Details

Troubleshooting and Malfunctions

T310(H) Model Troubleshooting

Specific troubleshooting steps and possible causes for the T310(H) model.

T420(H) and LC20 H Model Troubleshooting

Specific troubleshooting steps for T420(H) and LC20H models.

T460-T700 H and LC30-LC130 H Model Troubleshooting

Specific troubleshooting for T460-T700H and LC30-LC130H models.

T780-T1060 H Model Troubleshooting

Specific troubleshooting steps for T780-T1060H models.

How to Open the Unit

Faulty Operating and Switch Elements

Replacing the Timer Component

Detailed steps for replacing the timer component in the unit.

Replacing the Heating Switch

Instructions for removing and installing the heating switch.

Replacing the Control Thermostat

Steps to replace the control thermostat on specific models.

Replacing Overheating Protection (T310;H)

Procedure to replace the overheating protection component on T310/H models.

Replacing Overheating Protection (LC Models)

Procedure to remove and replace overheating protection on LC models.

Faulty PXE Transducer Disks

Assembling the Transducer System

Detailed steps for reassembling the transducer system correctly.

How to Replace the Transducer Tank

How to Replace the Ball Valve

How to Replace the Interference Filter

How to Replace the Ultrasonic Generator

Generator Repair and Replacement Procedures

General and detailed steps for repairing or replacing the ultrasonic generator.

Generator Replacement: T310-T420 H, LC20 H-LC30 H

Specific instructions for generator replacement on these models.

Generator Replacement: T460-T470 H, LC60 H

Specific instructions for generator replacement on these models.

Generator Replacement: TS540-T700 H, LC130 H

Steps for replacing components on the generator print for these models.

How to Replace Components on the Generator Print

Transistor Replacement (T310-T420 H, LC20 H-LC30 H)

Detailed steps for replacing transistors on the generator print for these models.

Transistor Replacement (T460-T470 H, LC60 H)

Detailed steps for replacing transistors on the generator print for these models.

Component Replacement (TS540-T700 H, LC130 H)

Steps for replacing components on the generator print for these models.

Component Replacement (T780-T1060 H)

Steps for replacing components on the generator print for these models.

Adjustment of the Generators

Generator Adjustment: T310-T420 H, LC20 H-LC30 H

Specific adjustment procedures for T310-T420H and LC20H-LC30H models.

Generator Adjustment: T460-T470 H, LC60 H

Specific adjustment procedures for T460-T470H and LC60H models.

Generator Adjustment: TS540-T700 H, LC130 H

Specific adjustment procedures for TS540-T700H and LC130H models.

Generator Adjustment: T780-T1060 H

Specific adjustment procedures for T780-T1060H models.

Putting into Operation and Trial Run