    Why does Elma T470H have no ultrasound but heating is OK?
    • J
      Joseph RobinsonAug 7, 2025
      If there's no ultrasound but the heating is working, the issue could be with the interference filter coil or the timer. In such cases, consider replacing the component or the complete interference filter.
  • D
    Dennis WillisAug 12, 2025
    What to do if Elma T470H has no ultrasound and no heating?
    • N
      ncampbellAug 12, 2025
      If there is no ultrasound and no heating, the issue might be a faulty printed board (BZW diode; transistors; diodes mains and HF). You can try replacing the component or the complete printed board. Alternatively, disconnect the heating cable and check the unit again with new fuses. If the short circuit doesn't occur again, the heating is faulty, and you should replace it.
  • K
    Kathryn PhillipsAug 17, 2025
    What to do if Elma Other fuse is released, with no heating and no ultrasound?
    • J
      Jonathan FuentesAug 17, 2025
      If the fuse is released, and there's no heating or ultrasound, it indicates a short circuit in the heating system. Disconnect the heating cable and check the unit again using new fuses. If the short circuit does not occur again, it confirms that the heating is faulty. Replace the faulty heating.
  • C
    Cindy Fisher DDSAug 22, 2025
    Why does Elma T470H Other have no ultrasound or weak ultrasound, but heating is OK?
    • Y
      ycookAug 22, 2025
      If there is no ultrasound or weak ultrasound, but the heating is okay, the issue could be with the output transformer, the output capacitor (6.8nF), or scorching in the transducer system. You may need to replace the component or the complete printed board, or open and check the transducer system.
  • A
    Andrea MatthewsAug 28, 2025
    What to do if Elma Other has no ultrasound but heating is OK?
    • J
      Jeffrey GatesAug 28, 2025
      If there's no ultrasound but the heating is working, possible causes include a faulty printed board (BZW diode; transistors; diodes) or a short circuit in the heating. Disconnect the heating cable and check again. You might need to replace a component, the complete printed board, or the heating element.
  • V
    Victor MitchellSep 1, 2025
    What to do if Elma T470H Other has no heating but ultrasound is OK?
    • J
      Jacqueline LoganSep 1, 2025
      If there is no heating but the ultrasound is working, the issue could be a faulty heating switch. If that's the case, replace the component.

Summary

General Information and Safety

Important Safety Warnings

Critical safety precautions to prevent injury and equipment damage.

Important Operating Instructions

Description of Operating Elements

Troubleshooting and Malfunctions

T310(H) Malfunctions

Troubleshooting guide for model T310(H), listing malfunctions and remedies.

T420(H) and LC20 H Malfunctions

Troubleshooting guide for models T420(H) and LC20H, listing malfunctions and remedies.

T460-T700 H and LC30-LC130 H Malfunctions

Troubleshooting guide for models T460-T700H and LC30-LC130H.

T780-T1060 H Malfunctions

Troubleshooting guide for models T780-T1060H, listing malfunctions and remedies.

How to Open the Unit

Faulty Operating and Switch Elements

Replacing the Timer

Procedure for replacing the timer component in the unit.

Replacing the Heating Switch

Instructions for replacing the heating switch element.

Replacing Control Thermostat

Procedure for replacing the control thermostat on specific models.

Replacing Overheating Protection (T310;H)

Steps to replace the overheating protection on T310/H models.

How to Replace the Heating Element

Faulty PXE Transducer Disks

How to Replace the Transducer Tank

How to Replace the Ball Valve

How to Replace the Interference Filter

How to Replace the Ultrasonic Generator

Replacing Generator Print Components

Generator Adjustment Procedures

Adjusting T310-T420 H; LC20 H-LC30 H Generators

Specific adjustment procedure for T310-T420H and LC20H-LC30H generators.

Adjusting T460-T470 H; LC60 H Generators

Adjustment procedure for T460-T470H and LC60H generators.

Adjusting TS540-T700 H; LC130 H Generators

Adjustment procedure for TS540-T700H and LC130H generators.

Adjusting T780-T1060 H Generators

Specific adjustment procedure for T780-T1060H generators.

Putting into Operation and Trial Run

Elma T470H Specifications

General IconGeneral
BrandElma
ModelT470H
CategoryOther
LanguageEnglish