EasyManua.ls Logo

Elma TP690 - User Manual

Default Icon
109 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Elma GmbH & Co KG Kolpingstr. 1-7 D-78224 Singen Tel. +49(0)7731/882-0 Fax +49(0)7731-266 info@elma-ultrasonic.com
Service Manual
Transsonic Analog
Transsonic LC
Ultrasonic Cleaning Units
y english y
www.elma-ultrasonic.com

Table of Contents

Questions and Answers

  • M
    Mark WoodsAug 7, 2025
    Why does Elma Other have no ultrasound but heating is OK?
    • K
      kimberly76Aug 7, 2025
      If there's no ultrasound but the heating is working, the issue could be with the interference filter coil or the timer. In such cases, consider replacing the component or the complete interference filter.
  • M
    Micheal HarrisAug 12, 2025
    What to do if Elma TP690 has no ultrasound and no heating?
    • L
      Lisa MartinezAug 12, 2025
      If there is no ultrasound and no heating, the issue might be a faulty printed board (BZW diode; transistors; diodes mains and HF). You can try replacing the component or the complete printed board. Alternatively, disconnect the heating cable and check the unit again with new fuses. If the short circuit doesn't occur again, the heating is faulty, and you should replace it.
  • N
    Noah SnyderAug 17, 2025
    What to do if Elma TP690 Other fuse is released, with no heating and no ultrasound?
    • R
      Rebecca ButlerAug 17, 2025
      If the fuse is released, and there's no heating or ultrasound, it indicates a short circuit in the heating system. Disconnect the heating cable and check the unit again using new fuses. If the short circuit does not occur again, it confirms that the heating is faulty. Replace the faulty heating.
  • D
    dgoodwinAug 22, 2025
    Why does Elma TP690 have no ultrasound or weak ultrasound, but heating is OK?
    • J
      Joseph ParsonsAug 22, 2025
      If there is no ultrasound or weak ultrasound, but the heating is okay, the issue could be with the output transformer, the output capacitor (6.8nF), or scorching in the transducer system. You may need to replace the component or the complete printed board, or open and check the transducer system.
  • J
    Joshua PageAug 27, 2025
    What to do if Elma Other has no ultrasound but heating is OK?
    • D
      Daniel HillAug 27, 2025
      If there's no ultrasound but the heating is working, possible causes include a faulty printed board (BZW diode; transistors; diodes) or a short circuit in the heating. Disconnect the heating cable and check again. You might need to replace a component, the complete printed board, or the heating element.
  • R
    rhonda15Sep 1, 2025
    What to do if Elma Other has no heating but ultrasound is OK?
    • M
      Mary BrewerSep 1, 2025
      If there is no heating but the ultrasound is working, the issue could be a faulty heating switch. If that's the case, replace the component.

Summary

General

How to use the Service Manual

Instructions on navigating and utilizing the service manual effectively.

Important safety warnings

Essential safety precautions to prevent injury and equipment damage.

Testing equipment, tools and measuring instruments

List and description of necessary tools and equipment for repairs.

Trouble shooting; List of malfunctions

T310(H)

Specific troubleshooting steps for the T310(H) model.

T420(H); LC20 H

Troubleshooting guide for T420(H) and LC20H models.

T460 - T700 H; LC 30 - LC 130 H

Troubleshooting for T460-T700H and LC30-LC130H units.

Faulty operating and switch elements

How to replace the timer

Detailed steps for replacing the timer component.

How to replace the heating switch

Instructions for replacing the heating switch.

How to replace the control thermostat T420 H-T1060 H

Procedure for replacing the control thermostat on specific models.

How to replace the heating

Faulty PXE transducer disks

How to replace the transducer tank

How to replace the interference filter

Interference filter on T310,T420, LC20 and LC30

Specifics of interference filter on T310, T420, LC20, LC30 models.

How to replace the ultrasonic generator

How to replace components on the generator print

T310 - T420 H; LC20 H - LC30 H

Replacing components on the generator print for T310-T420H, LC20H-LC30H.

Adjustment of the generators

Putting into operation; Trial run

Elma TP690 Specifications

General IconGeneral
BrandElma
ModelTP690
CategoryOther
LanguageEnglish

Related product manuals