Elma GmbH & Co KG Kolpingstr. 1-7 D-78224 Singen Tel. +49(0)7731/882-0 Fax +49(0)7731-266 info@elma-ultrasonic.com
Service Manual
Transsonic Analog
Transsonic LC
Ultrasonic Cleaning Units
y english y
www.elma-ultrasonic.com

Table of Contents

Questions and Answers

  • E
    Elizabeth StevensAug 8, 2025
    Why does Elma T570 Other have no ultrasound but heating is OK?
    • G
      George MorenoAug 8, 2025
      If there's no ultrasound but the heating is working, the issue could be with the interference filter coil or the timer. In such cases, consider replacing the component or the complete interference filter.
  • K
    kerrangelicaAug 12, 2025
    What to do if Elma T570 Other has no ultrasound and no heating?
    • M
      Mario MayAug 12, 2025
      If there is no ultrasound and no heating, the issue might be a faulty printed board (BZW diode; transistors; diodes mains and HF). You can try replacing the component or the complete printed board. Alternatively, disconnect the heating cable and check the unit again with new fuses. If the short circuit doesn't occur again, the heating is faulty, and you should replace it.
  • J
    Jean WilkinsonAug 17, 2025
    What to do if Elma T570 Other fuse is released, with no heating and no ultrasound?
    • O
      ochungAug 17, 2025
      If the fuse is released, and there's no heating or ultrasound, it indicates a short circuit in the heating system. Disconnect the heating cable and check the unit again using new fuses. If the short circuit does not occur again, it confirms that the heating is faulty. Replace the faulty heating.
  • J
    Joseph AllenAug 22, 2025
    Why does Elma Other have no ultrasound or weak ultrasound, but heating is OK?
    • J
      Justin MoralesAug 22, 2025
      If there is no ultrasound or weak ultrasound, but the heating is okay, the issue could be with the output transformer, the output capacitor (6.8nF), or scorching in the transducer system. You may need to replace the component or the complete printed board, or open and check the transducer system.
  • V
    vincent53Aug 27, 2025
    What to do if Elma T570 Other has no ultrasound but heating is OK?
    • K
      kenneth94Aug 27, 2025
      If there's no ultrasound but the heating is working, possible causes include a faulty printed board (BZW diode; transistors; diodes) or a short circuit in the heating. Disconnect the heating cable and check again. You might need to replace a component, the complete printed board, or the heating element.
  • K
    Kimberly CabreraSep 1, 2025
    What to do if Elma Other has no heating but ultrasound is OK?
    • A
      Amanda BarnesSep 1, 2025
      If there is no heating but the ultrasound is working, the issue could be a faulty heating switch. If that's the case, replace the component.

Summary

1 General

1.4 Important safety warnings

Critical safety precautions for handling the equipment.

1.5 Testing equipment, tools and measuring instruments

List and description of necessary tools for repair and maintenance.

2 Important operating instructions

5 Technical details

6 Troubleshooting and Malfunctions

Unit and System Faults

Diagnosing issues with the unit, ultrasound, and heating systems.

6.1 T310(H) Troubleshooting

6.2 T420(H); LC20 H Troubleshooting

6.3 T460 - T 700 H; LC 30 - LC 130 H Troubleshooting

6.4 T780 - T1060 H Troubleshooting

7 How to open the unit

8 Faulty operating and switch elements

8.2 Replacing the Timer

8.3 Replacing the Heating Switch

8.4 Replacing Control Thermostat (T420 H-T1060 H)

8.5 Replacing Overheating Protection (T310;H)

8.6 Replacing Overheating Protection (LC 20 H-LC 130 H)

9 How to replace the heating

10 Faulty PXE Transducer Disks

Causes of Transducer Disk Damage

Reasons for damage to PXE transducer disks, including tears and scorchings.

Transducer Disk Troubleshooting

Steps to diagnose issues with PXE transducer disks without opening.

11 How to replace the transducer tank

12 How to replace the ball valve

13 Interference Filter

13.1 Filter on T310, T420, LC20, LC30

Details on the interference filter integrated into the generator PCB for specific models.

13.2 Filter on T460 - T700 H

13.3 Filter on T780 - T1040;H

14 How to replace the ultrasonic generator

Diagnosing Generator Faults

Steps to find the cause of generator failure, checking pilot lamps and primary faults.

Generator Breakdown Causes

Common reasons for ultrasonic generator failure, including humidity and voltage issues.

14.1 Generator Replacement: T310-T420 H; LC20 H-LC30 H

14.2 Generator Replacement: T460-T470 H; LC60 H

14.3 Generator Replacement: TS540-T700 H; LC130 H

14.4 Generator Replacement: T780-T1060 H

15 Replacing Components on Generator Print

15.2 Component Replacement: T460-T470 H; LC60 H

15.3 Component Replacement: TS540-T700 H; LC130 H

15.4 Component Replacement: T780-T1060 H

16 Generator Adjustment

16.1 Measuring Instrument Setup

16.2 Measuring Process Preparation

16.3 Power Consumption Values (Ultrasound)

16.4 Adjusting T310-T420 H; LC20 H-LC30 H Generators

16.5 Adjusting T460-T470 H; LC60 H Generators

16.6 Adjusting TS540-T700 H; LC130 H Generators

16.7 Adjusting T780-T1060 H Generators

17 Putting into Operation; Trial Run

17.4 Ultrasound Check

Elma T570 Specifications

General IconGeneral
BrandElma
ModelT570
CategoryOther
LanguageEnglish