Elma GmbH & Co KG Kolpingstr. 1-7 D-78224 Singen Tel. +49(0)7731/882-0 Fax +49(0)7731-266 info@elma-ultrasonic.com
Service Manual
Transsonic Analog
Transsonic LC
Ultrasonic Cleaning Units
y english y
www.elma-ultrasonic.com

Table of Contents

Questions and Answers

  • B
    Brian GarrisonAug 8, 2025
    Why does Elma T890 Other have no ultrasound but heating is OK?
    • K
      klinemargaretAug 8, 2025
      If there's no ultrasound but the heating is working, the issue could be with the interference filter coil or the timer. In such cases, consider replacing the component or the complete interference filter.
  • T
    tiffany54Aug 12, 2025
    What to do if Elma Other has no ultrasound and no heating?
    • K
      kimberlymartinezAug 12, 2025
      If there is no ultrasound and no heating, the issue might be a faulty printed board (BZW diode; transistors; diodes mains and HF). You can try replacing the component or the complete printed board. Alternatively, disconnect the heating cable and check the unit again with new fuses. If the short circuit doesn't occur again, the heating is faulty, and you should replace it.
  • S
    Scott FoxAug 17, 2025
    What to do if Elma Other fuse is released, with no heating and no ultrasound?
    • Q
      qcoxAug 17, 2025
      If the fuse is released, and there's no heating or ultrasound, it indicates a short circuit in the heating system. Disconnect the heating cable and check the unit again using new fuses. If the short circuit does not occur again, it confirms that the heating is faulty. Replace the faulty heating.
  • R
    Robert MartinezAug 22, 2025
    Why does Elma T890 have no ultrasound or weak ultrasound, but heating is OK?
    • S
      Shawn FigueroaAug 22, 2025
      If there is no ultrasound or weak ultrasound, but the heating is okay, the issue could be with the output transformer, the output capacitor (6.8nF), or scorching in the transducer system. You may need to replace the component or the complete printed board, or open and check the transducer system.
  • Y
    yunderwoodAug 27, 2025
    What to do if Elma T890 Other has no ultrasound but heating is OK?
    • K
      Katie StaffordAug 27, 2025
      If there's no ultrasound but the heating is working, possible causes include a faulty printed board (BZW diode; transistors; diodes) or a short circuit in the heating. Disconnect the heating cable and check again. You might need to replace a component, the complete printed board, or the heating element.
  • L
    lindamoralesSep 1, 2025
    What to do if Elma T890 Other has no heating but ultrasound is OK?
    • J
      Justin ReedSep 1, 2025
      If there is no heating but the ultrasound is working, the issue could be a faulty heating switch. If that's the case, replace the component.

Summary

General

Important safety warnings

Critical safety precautions to prevent hazards during operation and maintenance.

Trouble shooting; List of malfunctions

T310(H)

Troubleshooting guide for the T310(H) model, listing malfunctions and remedies.

T460 - T 700 H; LC 30 - LC 130 H

Troubleshooting for T460-T700H and LC30-LC130H models, with common faults and fixes.

Faulty operating and switch elements

How to replace the timer

Detailed steps for replacing the timer component in the unit.

How to replace the ultrasonic generator

Adjustment of the generators

Adjustment on T310 - T420 H; LC20 H - LC30 H

Specific adjustment procedures for T310-T420H and LC20H-LC30H generator models.

Putting into operation; Trial run

Elma T890 Specifications

General IconGeneral
BrandElma
ModelT890
CategoryOther
LanguageEnglish

Related product manuals