Elma GmbH & Co KG Kolpingstr. 1-7 D-78224 Singen Tel. +49(0)7731/882-0 Fax +49(0)7731-266 info@elma-ultrasonic.com
Service Manual
Transsonic Analog
Transsonic LC
Ultrasonic Cleaning Units
y english y
www.elma-ultrasonic.com

Table of Contents

Questions and Answers

  • D
    David SchneiderAug 3, 2025
    Why does Elma T310H Other have no ultrasound but heating is OK?
    • R
      Renee MendezAug 3, 2025
      If there's no ultrasound but the heating is working, the issue could be with the interference filter coil or the timer. In such cases, consider replacing the component or the complete interference filter.
  • T
    thomasjoyAug 7, 2025
    What to do if Elma Other has no ultrasound and no heating?
    • L
      Logan BradyAug 7, 2025
      If there is no ultrasound and no heating, the issue might be a faulty printed board (BZW diode; transistors; diodes mains and HF). You can try replacing the component or the complete printed board. Alternatively, disconnect the heating cable and check the unit again with new fuses. If the short circuit doesn't occur again, the heating is faulty, and you should replace it.
  • C
    corymartinAug 12, 2025
    What to do if Elma T310H fuse is released, with no heating and no ultrasound?
    • T
      Traci WalkerAug 12, 2025
      If the fuse is released, and there's no heating or ultrasound, it indicates a short circuit in the heating system. Disconnect the heating cable and check the unit again using new fuses. If the short circuit does not occur again, it confirms that the heating is faulty. Replace the faulty heating.
  • S
    Stephanie BurnettAug 17, 2025
    Why does Elma T310H Other have no ultrasound or weak ultrasound, but heating is OK?
    • B
      Brandon HallAug 17, 2025
      If there is no ultrasound or weak ultrasound, but the heating is okay, the issue could be with the output transformer, the output capacitor (6.8nF), or scorching in the transducer system. You may need to replace the component or the complete printed board, or open and check the transducer system.
  • S
    Sharon OrozcoAug 23, 2025
    What to do if Elma T310H Other has no ultrasound but heating is OK?
    • M
      michael02Aug 23, 2025
      If there's no ultrasound but the heating is working, possible causes include a faulty printed board (BZW diode; transistors; diodes) or a short circuit in the heating. Disconnect the heating cable and check again. You might need to replace a component, the complete printed board, or the heating element.
  • C
    Christina YuAug 28, 2025
    What to do if Elma T310H has no heating but ultrasound is OK?
    • C
      Chelsea WoodAug 28, 2025
      If there is no heating but the ultrasound is working, the issue could be a faulty heating switch. If that's the case, replace the component.

Summary

General Information and Safety

Service Staff Requirements

Qualifications and knowledge required for authorized service personnel.

Critical Safety Warnings

Critical safety precautions to prevent injury and damage during operation and repair.

Required Tools and Equipment

List of necessary tools and equipment for repair and testing procedures.

Operating Instructions and Setup

Operating Element Descriptions

Troubleshooting Malfunctions

T310(H) Model Issues

Troubleshooting guide for T310(H) model, listing malfunctions and remedies.

T420(H) and LC20 H Issues

Troubleshooting guide for T420(H) and LC20H models, listing malfunctions and remedies.

T460-T700 H and LC30-LC130 H Issues

Troubleshooting guide for T460-T700H and LC30-LC130H models.

T780-T1060 H Model Issues

Troubleshooting guide for T780-T1060H models, listing malfunctions and remedies.

Unit Disassembly Procedures

Removing the Bottom Plate

Step-by-step instructions for removing the bottom plate of the unit.

Faulty Component Replacement

Replacing the Timer Component

Detailed steps for replacing the timer component in the unit.

Replacing the Heating Switch

Instructions for replacing the heating switch component.

Replacing Control Thermostat

Procedure for replacing the control thermostat on specific models.

Replacing T310;H Overheating Protection

Steps for replacing the overheating protection on T310/H models.

Removing LC 20 H-LC130 H Overheating Protection

Procedure to remove overheating protection from LC 20H-LC 130H.

Heating Element Replacement

PXE Transducer Disk Issues

Opening the Transducer System

Procedure for accessing and opening the transducer system.

Assembling the Transducer System

Steps for correctly assembling the transducer system after repair.

Transducer Tank Replacement

Transducer Tank Replacement Steps

Detailed steps for replacing the transducer tank.

Ball Valve Replacement

Interference Filter Replacement

Ultrasonic Generator Replacement

Generator Replacement T310-T420 H; LC20 H-LC30 H

Procedure to replace ultrasonic generator on T310-T420H, LC20H-LC30H.

Generator Replacement T460-T470 H; LC60 H

Procedure to replace ultrasonic generator on T460-T470H, LC60H.

Generator Replacement TS540-T700 H; LC130 H

Procedure to replace ultrasonic generator on TS540-T700H, LC130H.

Generator Replacement T780-T1060 H

Procedure to replace ultrasonic generator on T780-T1060H.

Generator Print Component Replacement

Components for T310-T420 H; LC20 H-LC30 H

Replacing components on generator print for T310-T420H, LC20H-LC30H.

Generator Adjustment Procedures

Initial Operation and Trial Run

Elma T310H Specifications

General IconGeneral
BrandElma
ModelT310H
CategoryOther
LanguageEnglish