EasyManua.ls Logo

ELSEC 764 - Page 15

ELSEC 764
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
12
Notes:
· If there are problems with infrared communication in they can sometimes be
solved by selectingFile” andOptions” in RView and setting the Infrared Mode
toSocket” and clicking on OK.
· After the link is established (Connected” is shown on the 764 and the PC), if
transmission is interrupted (e.g. by blocking the path between the 764 and the
computer infrared receiver) transmission will re-start from where it left off only if
communication is re-established within 15 seconds. If communication is blocked
for longer than this then the link is broken and the process must be re-started from
step 1. If all the data is not transferred for any reason the procedure can easily be
repeated, though see the information on file names above.
· If power has been lost then the 764 may forget how many records it has saved and
no data will be transferred. It is possible to force the 764 to transmit data by
selecting File\Get Data” from RView, the user then has to specify how many
records are required and they are transmitted whether they are valid or not.
· If battery power gets low during transmission a warning message is displayed but
transmission continues for as long as possible but may not complete. If this
happens the battery should be replaced and the data retrieved using the method