10
The temperature range then increments to the next available range with
every press of the set button: -
Leave the display showing your chosen range and press select to set up
another channel.
3.4.2 Voltage/current channel setup
These channels may be configured for voltage or current. There are
typically 18 voltage ranges and 2 current ranges. Voltage/current
channels have an additional range called the EU (Engineering Units)
range where the user can select specific engineering units (e.g. m/s) and
a max/min scale (e.g. 0/50 m/s). The appropriate voltage/current range is
then selected from the existing hardware ranges (e.g. 0/1 V). In the above
example, 0/50 m/s = 0/1V and the display in METER mode will show
readings in m/s. All data in this channel will be stored and downloaded
using m/s as the unit. In the above example, the CHANNEL SETUP
function will alternate between the following displays: -
The EU range may only be set up using Darca (please see price list or
contact your distributor for details). Whenever a voltage/current channel
has not been set up, the initial range is always the default EU range and is
displayed as follows: -
Press select to change channel. For example, select channel 5: -