50
9. Squirrels with pH channels
Some Squirrels have pH inputs. Such inputs require individual two-point
calibration, and temperature compensation. The Squirrel supports two
types of temperature compensation - measured and pre-set.
9.1 Channel Selection
The pH channels have two ranges available. These are actually both the
same range; the difference is where the temperature compensation
comes from. Choose from the pre-set temperature (which can be altered
in the STAT function), or from the associated temperature channel (as
illustrated below). If the latter is chosen, one of the -50°C to +150°C
ranges of the temperature channel must also be selected otherwise the
logger will perceive this as an error and switch the range to use the pre-
set compensation. In this situation, the display will show ‘INVALID –
CHECK!’ to alert the user that this has happened.
It is worth noting that if the temperature channel is selected but no probe
is plugged in, then compensation will again be taken from the pre-set
value. In this case, the pH ‘range’ is unchanged; the temperature input
would still be used if it were connected.
9.2 Pre-setting the Temperature
If the temperature of all the pH probes is known to be constant (or if high
accuracy is not required) then it is not necessary to use a temperature
probe for compensating the pH probe. The temperature of the pH probe
can be set in the STAT function.
Press and hold set to change the value. The longer set is pressed the
faster the value changes.