19
Guidance and manufacturer’s declaration
– electromagnetic emissions
Elvie Stride 2 uses RF energy
only for its internal function.
Therefore, its RF emissions
are very low and are not likely
to cause any interference in
nearby electronic equipment.
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Not Applicable
Group 1
Class B
Not Applicable
Emission Tests Compliance
Electromagnetic environmental
guidance
Elvie Stride 2 is suitable for use
in all establishments, including
domestic establishments and
those directly connected to the
public low voltage power supply
network that supplies buildings
used for domesticpurposes.
WARNING: This device should not be used adjacent to or stacked with other
equipment. And if adjacent or stacked use is necessary, this device should be
observed to verify normal operation in the configuration in which it will be used.
Elvie Stride 2 is intended for use in a Home Healthcare Environment. Elvie Stride 2
breast pump product is suitable for use in a specific electromagnetic environment
outlined below. The user of this product should assure that it is used in such an
environment.
Guidance and manufacturer’s declaration – electromagnetic immunity
Elvie Stride 2 breast pump product is suitable for use in a specific electromagnetic
environment outlined below. The user of this product should assure that it is used
in such an environment. Elvie Stride 2 breast pump was tested for immunity to
electromagnetic disturbances and passed using the following criteria:
1) No noticeable change in the operation of the breast pump signifying maximum
vacuum pressure is exceeded (-300 mmHg) or cycle speeds are degraded.
2) During charging, the pump did not turn on and start operating.
Electrostatic
discharge
(ESD) IEC
61000-4-2
+/- 2kV, +/- 4kV,
+/- 6kV, +/- 8kV
contact discharge
+/- 2kV, +/- 4kV,
+/- 6kV, +/- 8kV,
+/- 15 kV air
discharge
+/- 2kV, +/- 4kV,
+/- 6kV, +/- 8kV
contact discharge
+/- 2kV, +/- 4kV,
+/- 6kV, +/- 8kV,
+/- 15 kV
air discharge
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least30%.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m,
50 or 60 Hz
30 A/m,
50 or 60 H
It may be necessary to
position the breast pump
further from sources
of power frequency
magnetic fields or
to install magnetic
shielding. The power
frequency magnetic field
should be measured in
the intended installation
location to assure that it
is sucientlylow.
Immunity test
IEC 60601 test level Compliance level Electromagnetic
environment guidance