Control Solution Trouble Shooting
If your control solution testing is out of range (too high or too low), it
maybe caused by the following:
Possible Causes What you can do…..
Embrace N0 Code
Control Solution Testing
27
Wrong brand of Control Solu-
tion being used
Control solution not at room
temperature
Expired or contaminated con-
trol solution
Damaged test strip.
Meter malfunction
Make sure you are using
Embrace™ Control Solution.
Make sure the testing environ-
ment is between 50˚F~104˚F.
Check the expiration and open
bottle date on both the control
solution and test strips. Repeat
testing using a new test strip. If
the result is still out of range,
use a new bottle of control solu-
tion and retest.
If problem persists, call our au-
thorized dealer