EasyManua.ls Logo

embrace TALK - Conducting a Blood Glucose Test

embrace TALK
49 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Blood Glucose Monitoring System
30
Preparing Your Lancing Device
Embrace Lancing Device included in Starter Kit only (see page 4).
Step 1:
Twist off the adjustable cap counter clock wise and
insert a clean unused lancet as shown.
Step 2:
Twist the protective cover off the lancet exposing
the needle. Twist clockwise adjustable cover back
onto the lancing device.
Step 3.
Adjust the puncture setting on the adjustable cap
for the puncture depth level (1 is the lightest and 5
is the deepest).
Step 4:
With one hand holding middle way down the
lancing device, pull and release the opposite end of
the inserted lancet (you should hear a click sound).
The lancing device is now ready for pricking.
IMPORTANT:
Use a new sterile lancet every time you test to avoid cross
contamination. If alcohol wipes are used to cleanse the ngers,
make sure the area is dry before the blood sample is obtained. We
recommend washing with soap & warm water. Alcohol and Hand
Sanitizer may alter the blood test.