157-92, CSA Std C22.2 No. 213-M1987, ANSI-
ISA-12.27.01-2003
Markings:
Explosionproof for Class I, Division 1, Groups B, C and D; Class
II, Groups E, F, and G; Class III; Intrinsically Safe Class I,
Division 1 when connected in accordance with Rosemount
drawing 02088-1024, Temperature Code T3C; Ex ia; Class I
Division 2 Groups A, B, C and D; Type 4X; Factory Sealed
8.1.4 Europe
ED ATEX Flameproof
Certificate:
KEMA97ATEX2378X
Standards:
EN60079-0:2012 + A11:2013, EN60079-1:2014,
EN60079-26:2015
Markings:
II 1/2 G Ex db IIC T6…T4 Ga/Gb, T6(–60 °C ≤ T
a
≤ +70 °C),
T4/T5(–60 °C ≤ T
a
≤+80 °C);
Special Conditions for Safe Use (X):
1. This device contains a thin wall diaphragm. Installation, maintenance
and use shall take into account the environmental conditions to
which the diaphragm will be subjected. The manufacturer's
instructions for installation and maintenance shall be followed in
detail to assure safety during its expected lifetime.
2. Flameproof joints are not intended for repair.
3. Non-standard paint options may cause risk from electrostatic
discharge. Avoid installations that could cause electrostatic build-up
on painted surfaces, and only clean the painted surfaces with a damp
cloth. If paint is ordered through a special option code, contact the
manufacturer for more information.
Table 8-1: Process Connection Temperature
Temperature class Process connection
temperature
Ambient temperature
T6 -60 °C to +70 °C -60 °C to +70 °C
T5 -60 °C to +80 °C -60 °C to +80 °C
T4 -60 °C to +120 °C -60 °C to +80 °C
I1 ATEX Intrinsic Safety
Certificate:
BAS00ATEX1166X
Standards:
EN60079-0:2012+A11:2013, EN60079-11:2012