Figure 1-1: MeterLink download and registration
Select the MeterLink software and firmware bundle appropriate for your meter.
Complete the Online registration form and you will receive a confirmation email
with a hyperlink directing you to the download site.
NOTICE
After the download, follow the instructions in the Readme file. Do not attempt to
unzip the zipped firmware file. MeterLink unzips the compressed file using the
Tools → Program Download utility.
Refer to the MeterLink Software for Gas and Liquid Ultrasonic Meters Quick Start
Manual (00809-0100-7630) for installation instructions and to setup initial
communications. You may download the manual from the MeterLink web page:
www.emerson.com/meterlink
1.6 Rosemount 3814 Liquid Ultrasonic Flow
meter design
The Rosemount 3814 Liquid Ultrasonic Flow Meter is a four-path (eight
transducers) in-line meter designed to measure the difference in signal transit
time with and against the flow across one or more measurement path(s). A signal
transmitted in the flow direction travels faster than one transmitted against the
flow direction. Each measurement path is defined by a transducer pair in which
each transducer alternately acts as transmitter and receiver. The meter uses
transit time measurements and transducer location information to calculate the
mean velocity. Computer simulations of various velocity profiles demonstrate that
multiple measurement paths provide an optimum solution for measuring
asymmetric flow. The Rosemount 3814 Liquid Ultrasonic Flow Meter utilizes four
cross-bore, parallel-plane measurement paths, offers a high degree of
repeatability, bi-directional measurement and superior low-flow capabilities
without the compromises associated with conventional technologies.
Introduction Installation manual
January 2023 00825-0100-3814
10 Emerson.com/Rosemount