EasyManua.ls Logo

Emerson Rosemount 5300 Series - I2 INMETRO Intrinsic Safety; IB INMETRO FISCO

Emerson Rosemount 5300 Series
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ui Ii Pi Ci Li
Entity parameters
HART
30 V
dc
130 mA 1.0 W 7.26 nF Negligible
Entity parameters
Fieldbus
30 V
dc
300 mA 1.5 W 4.95 nF Negligible
9.9.3 IB INMETRO FISCO
Certificate
UL-BR 17.0188X
Standards
ABNT NBR IEC 60079-0:2013, ABNT NBR IEC 60079-11:2013,
ABNT NBR IEC 60079-26:2016, ABNT NBR IEC 60079-31:2014
Markings
Ex ia IIC T4 Ga (- 55 °C ≤ T
amb
≤ +60 °C)
Ex ia/ib IIC T4 Ga/Gb (- 55 °C ≤ T
amb
≤ +60 °C)
Ex ia IIIC T69 °C Da (- 50 °C ≤ T
amb
≤ +60 °C)
Ex ia/ib IIIC T69 °C Da/Db (- 50 °C ≤ T
amb
≤ +60 °C)
Special Conditions for Safe Use (X):
1. See certificate for Specific Conditions.
Ui Ii Pi Ci Li
FISCO parameters 17.5 V
dc
380 mA 5.32 W 4.95 nF <1 µH
9.10 China
9.10.1 E3 China Flameproof
Certificate
GYJ16.1095X
Standards
GB 3836.1/2/4/20-2010, GB 12476.1/5-2013, GB
12476.4-2010
Markings
Ex d ia IIC T4 Ga/Gb (-40 °C ≤ Ta ≤ +60 °C/+70 °C)
Ex tD A20 IP 66/67 T69 °C /T79 °C (-40 °C ≤ Ta ≤ +60 °C/+70 °C)
Special Conditions for Safe Use (X):
1. See certificate for Specific Conditions.
9.10.2 I3 China Intrinsic Safety
Certificate
GYJ16.1095X
Standards
GB 3836.1/2/4/20-2010, GB 12476.1/5-2013, GB
12476.4-2010