Ui Ii Pi Ci Li
Entity parameters
HART
30 V
dc
130 mA 1.0 W 7.26 nF Negligible
Entity parameters
Fieldbus
30 V
dc
300 mA 1.5 W 4.95 nF Negligible
9.9.3 IB INMETRO FISCO
Certificate
UL-BR 17.0188X
Standards
ABNT NBR IEC 60079-0:2013, ABNT NBR IEC 60079-11:2013,
ABNT NBR IEC 60079-26:2016, ABNT NBR IEC 60079-31:2014
Markings
Ex ia IIC T4 Ga (- 55 °C ≤ T
amb
≤ +60 °C)
Ex ia/ib IIC T4 Ga/Gb (- 55 °C ≤ T
amb
≤ +60 °C)
Ex ia IIIC T69 °C Da (- 50 °C ≤ T
amb
≤ +60 °C)
Ex ia/ib IIIC T69 °C Da/Db (- 50 °C ≤ T
amb
≤ +60 °C)
Special Conditions for Safe Use (X):
1. See certificate for Specific Conditions.
Ui Ii Pi Ci Li
FISCO parameters 17.5 V
dc
380 mA 5.32 W 4.95 nF <1 µH
9.10 China
9.10.1 E3 China Flameproof
Certificate
GYJ16.1095X
Standards
GB 3836.1/2/4/20-2010, GB 12476.1/5-2013, GB
12476.4-2010
Markings
Ex d ia IIC T4 Ga/Gb (-40 °C ≤ Ta ≤ +60 °C/+70 °C)
Ex tD A20 IP 66/67 T69 °C /T79 °C (-40 °C ≤ Ta ≤ +60 °C/+70 °C)
Special Conditions for Safe Use (X):
1. See certificate for Specific Conditions.
9.10.2 I3 China Intrinsic Safety
Certificate
GYJ16.1095X
Standards
GB 3836.1/2/4/20-2010, GB 12476.1/5-2013, GB
12476.4-2010