2. When the equipment is installed, particular precautions
must be taken to ensure, taking into account the effect of
process fluid temperature, that the ambient temperature
of the electrical housing of the equipment meets the
marked protection type temperature range.
ATEX Dust Certification
• EN 60079-0: 2012 + A11: 2013
• EN 60079-31: 2014
ND
Certificate: BaseefaATEX17.0019X;
II 2 D Ex tb IIIC T85°C Db (-20 °C ≤ Ta ≤ 70 °C)
Maximum Working Voltage = 42 VDC
Special conditions for safe use (X):
1. The enclosure may be made from aluminum alloy with a
protective polyurethane paint finish. The polyurethane
paint finish may constitute and electrostatic hazard and
must only be cleaned with a damp cloth.
2. When the equipment is installed, particular precautions
must be taken to ensure, taking into account the effect of
process fluid temperature, that the ambient temperature
of the electrical housing of the equipment meets the
marked protection type temperature range.
ATEX flameproof certification
• EN 60079-0: 2012 + A11: 2013
• EN 60079-1: 2014
• EN 60079-11: 2012
• EN 60079-26: 2015
E1
Certification No. DEKRA12ATEX0189X
Integral transmitter marked: ATEX marking:
II 1/2 G Ex db [ia] IIC T6...T2 Ga/Gb
Remote transmitter marked: ATEX marking:
II 2(1) G Ex db [ia Ga] IIC T6 Gb
Remote sensor marked: ATEX marking: II 1 G; Ex ia IIC T6...T2 Ga
EPL Ga piezo sensor and thermocouple connections.
EPL Gb transmitter enclosure.
Ambient temperature range: -50 °C ≤ Ta ≤ 70 °C
Maximum Working Voltage = 42 Vdc
April 2018 Quick Start Guide