Installation
Upon receipt, check for any visible damage which may have occurred in
transit. If any signs of damage are found then retain all packing material and
inform, within two working days, the carrier and the Company or its agent
from whom the unit was purchased.
The Primo Combination 860 operates at 18Vdc and if mains driven must only
be used with an EMS Physio SLA9000 power supply (as supplied with the
unit) which is connected to a mains supply of 100-240V ac. A power cord
appropriately rated/approved for the country of use must be used.
The SLA9000 power supply must only be connected to a mains supply with
a protective earth conductor. If the integrity of the earth connection is in
doubt, do not connect it to the mains supply (risk of electric shock with type
B applied parts). The unit must not be positioned in such a way that the mains
plug cannot easily be unplugged as the mains plug is the main disconnect
device.
The Primo Combination 860 unit is supplied with a large ultrasound
transducer and four medium-sized electrotherapy electrodes with their
associated patient lead. An optional small transducer is also available.
Plug the ultrasound transducer into the output socket on the front right of the
unit and the patient lead into the one to its left and connect the electrodes to
the yellow and blue cables.
Be careful not to subject the ultrasound transducers to rough handling such
as dropping onto a hard surface as this may impair performance.
Operation of the unit in close proximity (less than 1 metre) to shortwave
therapy equipment or radio-frequency mobile communication equipment
could result in the stimulation output being affected.
Permissible Environmental Conditions Of Use:
Temperature +10 to +35°C
Relative humidity 10 to 80%
Atmospheric pressure 500 to 1060hPa