Maintenance
The ultrasound transducers, electrodes and covers may be disinfected using
a 70% v/v aqueous solution of isopropyl alcohol. They are NOT suitable for
steam sterilisation or for disinfectants containing sodium hypochlorite.
N.B. Isopropyl alcohol is flammable and should be kept away from naked
flames. Isopropyl alcohol must not be brought into contact with eyes or
mouth.
The unit may be cleaned by wiping over with a damp cloth. The use of
abrasive materials and cleaning solvents should be avoided.
Regularly (at least monthly) inspect all treatment leads, cables and
connectors for signs of damage, particularly for cracks in the transducer
casing which may allow ingress of moisture. The ultrasonic output power
should be checked at least annually and a service calibration performed if
found to be out of specification.
There are no user serviceable parts inside the unit and it should not be
opened.
Full servicing instructions are available on request.