EasyManua.ls Logo

EMS Physio EMS860 - Technical Specifications and EMC Tests

EMS Physio EMS860
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OM860EN Iss 13
60
Appendix B - Technical specification
General
Power input (SLA9000) 100-240V ac 1.5A 50-60Hz
(EMS860) 18V, 3.33A (from external PSU SLA9000)
Classification (EN60601-1) Class 1, Type BF
Fuse Internal T5A
Size (h x w x d) 108 x 237 x 333 mm
Weight 1.3 kg
Treatment programs 48 user-defined set-ups.
Dose algorithm provided
Ultrasound
Frequency 1.1 MHz ±5% and 3.4 MHz ±5%
Maximum intensity 1.5 W/cm
2
in CW
3.0 W/cm
2
in pulsed modes
Maximum output power 6 W average
Output modes CW and pulsed 1:1, 1:2, 1:3, 1:4 and 1:9
Pulse duration 2 ms
Treatment timer 0 to 30 minutes (treatment linked)
Contact monitor Light on transducer
Large Ultrasound Transducer
ERA 4 cm
2
(IEC 61689)
5 cm
2
(21 CFR 1050.10)
BNR <5
Beam type Collimated
Small Ultrasound Transducer
1MHz 3MHz
ERA 0.6 cm
2
0.4 cm
2
BNR <5 <5
Beam Type Divergent Collimated
Transducers for use with the Primo Combination 860, Primo Therasonic 360
and 460 are fully interchangeable and suitable for underwater treatment
(IPx7 rated).