OM860EN Iss 13
61
Pulse Off Duty Temporal peak
Mode Frequency Time Cycle to average ratio
1:1 250 Hz 2 ms 50% 2:1
1:2 166 Hz 4 ms 33% 3:1
1:3 125 Hz 6 ms 25% 4:1
1:4 100 Hz 8 ms 20% 5:1
1:9 50 Hz 8 ms 10% 10:1
The pulse width is fixed at 2 ms
Stimulation Constant Current (CC) measured into 500 Ohm, Constant
Voltage (CV) into open circuit. Operation outside a safe range is prevented
by the electrode fault monitor alarm.
Interferential 4-pole
Carrier frequency 2 kHz, 4 kHz or 8 kHz
AMF 0 250 Hz in 1 Hz increments
Swing pattern 1|1, 6|6 or 6^6
Vector 10s, 20% both channels
Output type CC 0-100mA peak
CV 0-70V peak
Interferential 2-pole
Carrier frequency 2 kHz, 4 kHz or 8 kHz
AMF 0 250 Hz in 1 Hz increments
Swing pattern 1|1, 6|6 or 6^6
Output type CC 0-100mA peak
CV 0-70V peak