User Programs
The Interferential 960/Multidyne 970 can store up to 16 user defined set-
ups. Touch the User Programs button in the Home screen or in any
stimulation set up screen.
A screen will appear with a scrollable list of program slots the active one
is highlighted in the middle. Touching ‘Save’ will store the settings from the
last open set up screen to this slot if the slot isn’t empty a pop-up window
will appear asking you to confirm or cancel the save process (to prevent
unintentional over-writing of a previously saved program).
To recall a program simply touch its program slot button and a set up
screen will open showing the previously stored parameters.
The ‘Notesbutton in any user program set-up screen opens a qwerty style
keypad that allows you to save memo information about the program the
first 30 characters recorded will be displayed as the title of that user
program.