1. Setting up
Setting Up
Electrical connections
* The voltage and frequency for the Octagon 200CL is
given on the rating label.
* Ensure that the values agree with the existing power
supply.
* Connect the Octagon 200CL to the power supply using
the connection cable provided.
* When connecting the power cable to the mains
external fusing is necessary according to the
regulations of the installation location
Do not connect to any other supply other than stated on the rating label,
otherwise electrical and mechanical components can be damaged.
Ambient temperature: 5°C to 40°C
If the ambient temperature is exceeded or drops below the specified value
the electrical and mechanical components may become damaged and the
performance data can change to an unknown degree.
Atmospheric humidity:
Maximum relative humidity 80% at temperatures up to 31°C, with linear
reduction down to 50% relative humidity at 40°C.
At high atmospheric humidity the electrical and mechanical components
may become damaged and the performance data can change to an
unknown degree.