iTHERM TM412
Endress+Hauser 27
TA30R (high version for two transmitters) Specification
116 (4.57)
64 (2.52)
64 (2.52)
96 (3.8)
25
(1)
A0034644
• Degree of protection: IP69K (NEMA Type 4x encl.)
• Temperature: –50 to +130 °C (–58 to +266 °F) without cable
gland
• Material: stainless steel 316L, abrasive-blasted or polished
Seals: EPDM
• Cable entry thread ½" NPT and M20x1.5
• Weight: 460 g (16.23 oz)
• For two head transmitter
• Protection armature connection: M24x1.5 or ½" NPT
• Ground terminal: internal in standard version; external
terminal optionally available
• Available in conjunction with 3-A marked sensors
TA30S Specification
122 (4.8)
28
(1.1)
78 (3.1)
50 (1.97)
87 (3.43)
81.7 (3.22)
A0017146
• Degree of protection: IP65 (NEMA Type 4x encl.)
• Temperature: –40 to +85 °C (–40 to +185 °F) without cable
gland
• Material: polypropylene (PP), FDA-compliant, seals: O-ring
EPDM
• Cable entry thread: ¾" NPT (with adapter for ½" NPT),
M20x1.5
• Protective assembly connection: ½" NPT
• Color: white
• Weight: approx. 100 g (3.5 oz)
• Ground terminal: only internal via auxiliary terminal
• Available in conjunction with 3-A marked sensors