EasyManua.ls Logo

Endress+Hauser iTHERM TM411 - Page 26

Endress+Hauser iTHERM TM411
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
iTHERM TM412
26 Endress+Hauser
TA30P Specification
!83 (3.3)
114 (4.5)
85 (3.4)
41.5 (1.63)
28
(1.1)
A0023477
Protection class: IP65
Max. temperature: –40 to +120 °C (–40 to +248 °F)
Material: polyamide (PA), antistatic
Seals: silicone
Threaded cable entry: ½" NPT and M20x1.5
Protection armature connection: M24x1.5 or ½" NPT
Two head transmitters can be mounted. In the standard
version, one transmitter is mounted in the terminal head
cover and an additional terminal block is installed directly on
the insert.
Head and cap color: black
Weight: 135 g (4.8 oz)
Types of protection for use in hazardous locations: Intrinsic
Safety (G Ex ia)
Ground terminal: only internal via auxiliary clamp
TA30R (optionally with display window in
cover)
Specification
64 (2.52)
96 (3.8)
64 (2.52)
71 (2.8)
25 (1)
96 (3.8)*
A0017145
* Dimensions of version with display window
in cover
Degree of protection - standard version: IP69K (NEMA Type
4x encl.)
Degree of protection - version with display window: IP66/68
(NEMA Type 4x encl.)
Temperature: –50 to +130 °C (–58 to +266 °F) without cable
gland
Material: stainless steel 316L, abrasive-blasted or polished
Seals: EPDM
Display window: polycarbonate (PC)
Cable entry thread ½" NPT and M20x1.5
Weight
Standard version: 360 g (12.7 oz)
Version with display window: 460 g (16.23 oz)
Display window in cover optionally for head transmitter with
display TID10
Protection armature connection: M24x1.5 or ½" NPT
Ground terminal: internal in standard version; external
terminal optionally available
Available in conjunction with 3-A marked sensors

Related product manuals