Electrical connection Liquistation CSF34
38 Endress+Hauser
5.4 Connecting additional inputs, outputs or relays
L
WARNING
Module not covered
No shock protection. Danger of electric shock!
Change or extend the hardware: always fill the slots from top to bottom. Do not leave any
gaps.
If all of the slots are not occupied: always insert a dummy cover or end cover in the slot
below the last module. This ensures that the unit is shock-protected.
Always ensure shock protection is guaranteed particularly in the case of relay modules (2R,
4R, AOR).
If additional shields are required, connect them with PE centrally in the control cabinet via
terminal blocks supplied by the customer.
5.4.1 Digital inputs and outputs
DIO module
 29 Module
47
47
48
48
+
+
1
2
1
2
1
2
+
+
+
+
45
46
45
46
91
92
91
92
 30 Wiring
diagram
A maximum of 2 optional DIO modules are supported