EasyManua.ls Logo

Endress+Hauser Micropilot FMR50 - Performance Characteristics; Reference Operating Conditions; Influence of Ambient Temperature

Endress+Hauser Micropilot FMR50
96 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR50
Endress+Hauser 33
Performance characteristics
Reference operating
conditions
Temperature = +24 °C (+75 °F) ±5 °C (±9 °F)
Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi)
Humidity = 60 % ±15 %
Reflector: metal plate with a diameter ≥ 1 m (40 in)
No major interference reflections inside the signal beam
Maximum measured error
Typical data under reference operating conditions: DIN EN IEC 61298-2 / DIN EN IEC 60770-1;
percentage values in relation to the span.
Device version Value Output
digital analog
1)
Standard Sum of non-linearity, non-repeatability
and hysteresis
± 2 mm (0.08 in) ± 0.02 %
Offset/Zero ± 4 mm (0.2 in) ± 0.03 %
Version with the "Advanced
dynamics" application package
Sum of non-linearity, non-repeatability
and hysteresis
± 3 mm (0.12 in) ± 0.02 %
Offset/Zero ± 4 mm (0.2 in) ± 0.03 %
1) Only relevant for 4-20mA current output; add error of the analog value to the digital value
Differing values in near-range applications
2 (0.08)
0
-2 (-0.08)
A 1 (3.3)
4 (0.16)
-4 (-0.16)
R
D [mm] ([in])
D [m] ([ft])
A0019041
 18 Maximum measured error in near-range applications; values for standard version
Maximum measured error
A Lower edge of the antenna
D Distance from the lower edge A of the antenna
R Reference point of the distance measurement

Other manuals for Endress+Hauser Micropilot FMR50

Related product manuals