Micropilot FMR50
Endress+Hauser 41
Align the antenna so that it is perpendicular to the product surface.
Optionally, an adjustable flange seal (accessory) can be used for alignment
A marking is provided on the gland to aid the alignment. This marking must be aligned towards
the tank wall as much as possible.
90°
90°
90°
90°
90°
A0019439
Depending on the device version the marking may be a circle or two parallel lines.
Information concerning nozzles
H
øD
A0016868
 22 Nozzle diameter and height for horn antennas with slip-on flange
Nozzle diameter D Maximum nozzle height H
max
80 mm (3 in) 300 mm (11.8 in)
100 mm (4 in) 400 mm (15.8 in)
150 mm (6 in) 500 mm (19.7 in)