EasyManua.ls Logo

Endress+Hauser Micropilot FMR51 - Page 11

Endress+Hauser Micropilot FMR51
114 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Micropilot FMR51, FMR52
Endress+Hauser 11
Device Process tank with agitator
A0018837
Turbulent surface (e.g. filling from above, agitators, baffles)
Antenna size
FMR51 40 mm (1½ in) 50 mm (2 in) 80 mm (3 in) 100 mm (4 in)
FMR52 - 50 mm (2 in) 80 mm (3 in) -
B C D
1
(3.2)
2
(6.6)
3
(9.8)
5
(16)
A0018867
B C D
2
(6.6)
3
(9.8)
5
(16)
7.5
(25)
10
(32)
A0018868
B C D
2.5
(8.2)
5
(16)
12
(39)
15
(49)
8
(26)
A0018869
B C D
4
(13)
(16)
5
8
(26)
15
(49)
20
(65)
10
(32)
A0018870
Measuring range [m (ft)]
Legend
Measuring range of the standard version
Measuring range for the "Advanced dynamics" application package (product structure: feature 540:
"Application Package", Option EM: "Advanced dynamics")