Micropilot FMR51, FMR52
16 Endress+Hauser
Multidrop current 4.0 mA
Set-up time 15 s
PROFIBUS PA
Manufacturer ID 17 (0x11)
Ident number 0x1559
Profile version 3.02
GSD file Information and files under:
• www.endress.com
• www.profibus.org
GSD file version
Output values Analog Input:
• Level linearized
• Distance
• Terminal voltage
• Electronic temperature
• Absolute echo amplitude
• Relative echo amplitude
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
Digital Input:
• Digital output AD 1
• Digital output AD 2
• Switch output
Input values Analog Output:
• Analog value from PLC (for sensor block external pressure to compensate gas phase
effects)
• Analog value from PLC to be indicated on the display
Digital Output:
• Extended diagnostic block
• Level limiter
• Sensor block measurement on
• Sensor block save history on
• Status output
Supported functions • Identification & Maintenance
Einfachste Geräteidentifizierung seitens des Leitsystems und des Typenschildes
• Automatic Ident Number Adoption
GSD compatibility mode with respect to the preceding product Micropilot M FMR2xx
• Physical Layer Diagnostics
Installation check of the PRFIBUS segment and the Micropilot FMR5x via the terminal
voltage and telegram surveillance.
• PROFIBUS Up-/Download
Up to 10 times faster writing and reading of parameters via PROFIBUS up-/download
• Condensed Status
Simple and self-explanatory diagnostic information by categorization of occurring
diagnostic messages.
FOUNDATION Fieldbus
Manufacturer ID 0x452B48
Device type 0x1028
Device Revision 0x01
DD Revision Information and files can be found:
• www.endress.com
• www.fieldcommgroup.org
CFF Revision
Device Tester Version (ITK Version) 6.0.1
ITK Test Campaign Number IT085300
Link Master (LAS) capable yes