Micropilot FMR51, FMR52
36 Endress+Hauser
Installation
Installation conditions Mounting position
A
1 2 3
A0016882
Recommended distance A from wall to outer
edge of nozzle: ~ 1/6 of tank diameter.
Nevertheless the device should not be
installed closer than 15 cm (5.91 in) to the
tank wall.
Not in the center (2), as interference can
cause signal loss.
Not above the fill stream (3).
It is recommended to us a weather protection
cover (1) in order to protect the device from
direct sun or rain.
Vessel installations
A0018944
Avoid any installations (point level switches,
temperature sensors, braces, vacuum rings,
heating coils, baffles etc.) inside the signal
beam. Take into account the beam angle
→  38.