EasyManua.ls Logo

Endress+Hauser NRF81 - Incoming acceptance and product identification; Incoming Goods Inspection; Product Identification Methods

Endress+Hauser NRF81
36 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Tankside Monitor NRF81 Electrical connection
Endress+Hauser 13
Terminal area G (for High voltage AC power supply and Low voltage AC power supply)
G1: N
G2: not connected
G3: L
Terminal area G (for Low voltage DC power supply)
G1: L-
G2: not connected
G3: L+
Terminal area: Protective ground
Module: Protective ground connection (M4 screw)
A0018339
 5 Terminal area: Protective ground
Power supply
D
E
F
C
B
A
1
1
1
1 3
2
2 4
1
HR
CDI
WP
on
SIM
2
2
3
3
4
4
1
1
2
2
3
3
4
4
5
5
6
6
7
7
8
8
i
G
1
3
2
POWER
F
1 3
2 4
HR
WP
SIM
G
1
3
2
POWER
4
A0033413
G1 N
G2 not connected
G3 L
4 Green LED: indicates power supply
The supply voltage is also indicated on the nameplate.
Supply voltage
High voltage AC power supply:
Operational value:
100 to 240 V
AC
(- 15 % + 10 %) = 85 to 264 V
AC
, 50/60 Hz

Other manuals for Endress+Hauser NRF81