Cable entries
For electrical connection, the following cable entries are available:
EEx ib version (intrinsically safe)
Optional M20x1,5 or 1/2" NPT or G 1/2" threads or PG 13,5 cable glands.
For cable entry connections between sensor and transmitter E+H use U71, PG 13,5,
Typ 252221 1s04 of Pflitsch company.
EEx d version (explosion proof)
Only M20x1,5 or 1/2" NPT or G 1/2" threads.
Cable specification for the remote version
Maximum capacity of the connection cable is 270 nF/km.
Maximum inductivity of the connection cable 0,5 mH/km.
These values are fullfilled by the cable supplied by E+H (max. 30 meters).