RIA452 Technical data
Endress+Hauser 53
10.4.3 Material
• Housing front: ABS plastic, galvanized
• Housing casing: plastic PC10GF
10.4.4 Terminals
Pluggable screw terminals, core size 1.5 mm
2
(16 AWG) solid, 1.0 mm
2
(18 AWG) strand
with wire ferrule
10.5 Human interface
10.5.1 Display elements
G09-RIA452ZZ-07-01-00-xx-000
Fig. 28: LC display of panel meter
Item 1: Device status LEDs: green - device ready for operation; red - device or sensor malfunction
Item 2: Bar graph with overrange and underrange
Item 3: 7-digit 14-segment display
Item 4: Unit and text field 9x77 dot matrix
Item 5: Relay status display: if power is supplied to a relay, the symbol is displayed
Item 6: Status display, digital inputs
Item 7: Symbol for 'device operation blocked'
•Display range
-99999 to +99999 for measured values
0 to 9999999 for counter values
• Signaling
–Relay activation
– Measuring range overshoot/undershoot
10.5.2 Operating elements
Jog/shuttle dial
10.5.3 Remote operation
Configuration
The device can be configured with the PC software ReadWin
®
2000.
Interface
CDI interface at device; connection to PC via USB box (see 'Accessories')
RS232 interface at device; connection with serial interface cable (see 'Accessories')