BA02191C/66/EN/01.21
Operating Instruction
SS2100 H
2
S TDLAS Gas Analyzer
Class I, Division 2 Group A, B, C, D T3 / T3C
Class I, Zone 2 IIC T3 / T3C
Type 4X, IP66

Table of Contents

Other manuals for Endress+Hauser SS500

Questions and Answers

  • B
    Brian BallardAug 17, 2025
    What to do if Endress+Hauser Measuring Instruments have a power fail error?
    • S
      Scott VelazquezAug 17, 2025
      If your Endress+Hauser Measuring Instruments show a power fail error, turn off the power and check the optical head cables for loose connections. Also, inspect the inlet and outlet tubes to ensure they are not under stress. If instructed, clean the mirrors by following the provided instructions.
  • J
    Jessica BlakeAug 20, 2025
    How to fix electrical noise in Endress+Hauser Measuring Instruments?
    • C
      Colleen MolinaAug 20, 2025
      High levels of electrical noise can interfere with laser operation, causing instability in Endress+Hauser Measuring Instruments. To resolve this, ensure the analyzer is connected to a properly grounded power source.
  • M
    martinezmelanieAug 22, 2025
    What to do if the current loop is stuck at 4 mA or 20 mA on Endress+Hauser SS500 Measuring Instruments?
    • N
      noah77Aug 22, 2025
      If the current loop is stuck at 4 mA or 20 mA on your Endress+Hauser Measuring Instruments, it indicates that an alarm has been triggered. You should try to reset the alarm. If the problem persists, it is likely on the HC12 main board, and the unit should be returned to the factory for service.
  • L
    Luis CabreraAug 25, 2025
    What to do if Endress+Hauser SS500 shows a P/T Fail Error?
    • M
      Michelle CaseyAug 26, 2025
      If your Endress+Hauser Measuring Instruments display a P/T Fail Error, check that the actual pressure and temperature in the measurement cell are within the specified limits. Also, ensure the pressure/temperature cable is securely connected.
  • D
    dawn15Aug 27, 2025
    What to do if Endress+Hauser Measuring Instruments shows a Spectrum Fail Error?
    • M
      Michele KimAug 28, 2025
      If your Endress+Hauser Measuring Instruments display a Spectrum Fail Error, turn the analyzer off for 30 seconds and then turn it on again.
  • E
    Elizabeth GutierrezAug 30, 2025
    How to reduce electrical noise in Endress+Hauser SS500?
    • J
      jeremy76Aug 30, 2025
      High levels of electrical noise can interfere with laser operation, causing instability. To mitigate this, always connect the Endress+Hauser Measuring Instruments analyzer to a properly grounded power source.
  • U
    uedwardsSep 1, 2025
    What to do if I suspect gas leaks in Endress+Hauser SS500 Measuring Instruments?
    • R
      Rebecca Huynh DDSSep 1, 2025
      If you suspect gas leaks, especially after relocation, replacement, or factory service of your Endress+Hauser Measuring Instruments, periodically leak test the supply lines. Outside air leaking into the sample supply line can cause issues.
  • M
    Marie Weaver MDSep 4, 2025
    What to do if the current loop is stuck at 4 mA or 20 mA in Endress+Hauser Measuring Instruments?
    • L
      laura63Sep 4, 2025
      If the current loop is stuck, check the display for an error message on your Endress+Hauser Measuring Instruments. If an alarm has been triggered, reset the alarm.
  • M
    Mary SpencerSep 6, 2025
    Why am I not getting enough flow to the sample cell in my Endress+Hauser Measuring Instruments?
    • A
      amanda46Sep 6, 2025
      If you're not getting enough flow to the sample cell, check both the micro filter and membrane separator for contamination in your Endress+Hauser Measuring Instruments. Replace them if necessary.
  • M
    Mark LongSep 9, 2025
    What to do if the front panel display is not lit on Endress+Hauser Measuring Instruments?
    • E
      efranklinSep 9, 2025
      If the front panel display is not lit and no characters appear, check for correct voltage on the terminal block input of your Endress+Hauser Measuring Instruments. Observe polarity on DC powered units.

Summary

Safety Precautions for the SS2100 Analyzer

Potential Risks Affecting Personnel

Actions to take when faced with hazardous situations during analyzer service.

Explosion Hazard

Precautions for working in hazardous areas to avoid ignition sources.

Installation Procedures for the SS2100 Analyzer

Connecting Electrical Power to the Analyzer

Procedure for connecting the analyzer to power sources (120/240 VAC or 24 VDC).

Protective Chassis and Ground Connections

Requirements for connecting protective and chassis grounds for safety.

Sample Conditioning System (SCS) Operation

Starting up the SCS

Procedure for initial startup and warm-up of the Sample Conditioning System.

Appendix B: Maintenance and Troubleshooting

Gas Leaks

Common causes of erroneous measurements and recommended leak testing procedures.

Endress+Hauser SS500 Specifications

General IconGeneral
ManufacturerEndress+Hauser
Process ConnectionFlange
MaterialStainless Steel
Protection RatingIP66/IP67
ApprovalsATEX, IECEx
Process Temperature Range-20 to 50 °C