P/N 4900002285 rev B
SS500/SS2000
Single Channel Gas Analyzer
Hardware Installation
and Maintenance Manual

Table of Contents

Questions and Answers

  • B
    Bobby EllisAug 18, 2025
    What to do if Endress+Hauser SpectraSensors SS500 have a power fail error?
    • S
      Sharon BartonAug 18, 2025
      If your Endress+Hauser Measuring Instruments show a power fail error, turn off the power and check the optical head cables for loose connections. Also, inspect the inlet and outlet tubes to ensure they are not under stress. If instructed, clean the mirrors by following the provided instructions.
  • M
    michael21Aug 20, 2025
    How to fix electrical noise in Endress+Hauser SpectraSensors SS500 Measuring Instruments?
    • M
      Misty RobersonAug 21, 2025
      High levels of electrical noise can interfere with laser operation, causing instability in Endress+Hauser Measuring Instruments. To resolve this, ensure the analyzer is connected to a properly grounded power source.
  • A
    April GeorgeAug 23, 2025
    What to do if the current loop is stuck at 4 mA or 20 mA on Endress+Hauser Measuring Instruments?
    • M
      markstokesAug 23, 2025
      If the current loop is stuck at 4 mA or 20 mA on your Endress+Hauser Measuring Instruments, it indicates that an alarm has been triggered. You should try to reset the alarm. If the problem persists, it is likely on the HC12 main board, and the unit should be returned to the factory for service.
  • G
    Gloria ObrienAug 26, 2025
    What to do if Endress+Hauser Measuring Instruments shows a P/T Fail Error?
    • D
      dana10Aug 26, 2025
      If your Endress+Hauser Measuring Instruments display a P/T Fail Error, check that the actual pressure and temperature in the measurement cell are within the specified limits. Also, ensure the pressure/temperature cable is securely connected.
  • H
    Haley McclainAug 28, 2025
    What to do if Endress+Hauser SpectraSensors SS500 shows a Spectrum Fail Error?
    • S
      scottcisnerosAug 28, 2025
      If your Endress+Hauser Measuring Instruments display a Spectrum Fail Error, turn the analyzer off for 30 seconds and then turn it on again.
  • S
    Sandra KnoxAug 30, 2025
    How to reduce electrical noise in Endress+Hauser Measuring Instruments?
    • L
      Lisa EllisAug 30, 2025
      High levels of electrical noise can interfere with laser operation, causing instability. To mitigate this, always connect the Endress+Hauser Measuring Instruments analyzer to a properly grounded power source.
  • A
    Ashley TranSep 2, 2025
    What to do if I suspect gas leaks in Endress+Hauser SpectraSensors SS500?
    • A
      Abigail BeardSep 2, 2025
      If you suspect gas leaks, especially after relocation, replacement, or factory service of your Endress+Hauser Measuring Instruments, periodically leak test the supply lines. Outside air leaking into the sample supply line can cause issues.
  • W
    William MedinaSep 4, 2025
    What to do if the current loop is stuck at 4 mA or 20 mA in Endress+Hauser SpectraSensors SS500?
    • C
      Collin JordanSep 5, 2025
      If the current loop is stuck, check the display for an error message on your Endress+Hauser Measuring Instruments. If an alarm has been triggered, reset the alarm.
  • G
    Gary MartinezSep 7, 2025
    Why am I not getting enough flow to the sample cell in my Endress+Hauser SpectraSensors SS500 Measuring Instruments?
    • E
      erinboydSep 7, 2025
      If you're not getting enough flow to the sample cell, check both the micro filter and membrane separator for contamination in your Endress+Hauser Measuring Instruments. Replace them if necessary.
  • D
    Danielle LinSep 9, 2025
    What to do if the front panel display is not lit on Endress+Hauser SpectraSensors SS500?
    • K
      Kimberly JonesSep 10, 2025
      If the front panel display is not lit and no characters appear, check for correct voltage on the terminal block input of your Endress+Hauser Measuring Instruments. Observe polarity on DC powered units.

Endress+Hauser SpectraSensors SS500 Specifications

General IconGeneral
Measurement PrincipleTunable Diode Laser Absorption Spectroscopy (TDLAS)
Device TypeGas Analyzer
Measured GasCO2, CH4, H2S, NH3
Response TimeLess than 2 seconds
Operating Temperature-20 to +50 °C
Power Supply24 VDC
Communication ProtocolModbus, Ethernet
Outputs4-20 mA
Enclosure RatingNEMA 4X
ApplicationChemical, Petrochemical
MountingWall
Repeatability±0.5% of reading