SS2100r H
2
S Analyzer
(Rack Mount)
Hardware Manual
P/N 4900002189 rev D

Table of Contents

Questions and Answers

  • Q
    qsmithAug 3, 2025
    Why is electrical noise affecting my Endress+Hauser SpectraSensors SS2100r Measuring Instruments?
    • M
      Mr. Samuel DuncanAug 3, 2025
      High levels of electrical noise can interfere with laser operation of Endress+Hauser Measuring Instruments. To address this, always connect the analyzer to a properly grounded power source.
  • D
    David ChaseAug 9, 2025
    How to prevent gas leaks in Endress+Hauser SpectraSensors SS2100r?
    • D
      Dennis BradfordAug 9, 2025
      To prevent gas leaks in Endress+Hauser Measuring Instruments, periodically leak-test the supply lines, especially after relocation, replacement, or factory service with reconnection of supply lines. Also, avoid using plastic tubing for sample lines; instead, use 1/4” O.D. x 0.035” wall thickness, seamless stainless steel tubing.
  • M
    Mr. Christopher JenkinsAug 15, 2025
    What to do if Endress+Hauser SpectraSensors SS2100r are not operating at start up?
    • V
      Veronica JosephAug 15, 2025
      If your Endress+Hauser Measuring Instruments are not operating at start up, ensure that the power is connected to both the analyzer and the power source, and that the switch is turned on.
  • L
    leonard68Aug 21, 2025
    What to do if Endress+Hauser Measuring Instruments are not operating after start up?
    • S
      Selena WilliamsAug 21, 2025
      If your Endress+Hauser Measuring Instruments are not operating after start up, verify that the power source is adequate (120 VAC @ 50-60 Hz, 9-16 VDC, 18-32 VDC), referring to the Calibration Certificate for specific power requirements. Also, check the fuse(s) and replace any blown fuses with equivalent ones, referring to Table 1–1.
  • B
    Brittany MorrisNov 26, 2025
    How to deal with contamination in Endress+Hauser Measuring Instruments?
    • R
      russellwilsonNov 26, 2025
      Contamination in Endress+Hauser Measuring Instruments can be addressed by periodically cleaning the gas sampling lines due to long exposure to high humidity. Additionally, ensure that a membrane separator filter (included with most systems) is installed ahead of the analyzer and operating normally. To clean the sampling line turn off the sample valve at the tap in accordance with site lock-out, tag-out rules. Disconnect the gas sampling line from the sample supply port of the analyzer. Wash the sampling line with isopropyl alcohol or acetone and blow dry with mild pressure from a dry air or nitrogen source. Once the sampling line is completely free of solvent, reconnect the gas sampling line to the sample supply port of the analyzer. Check all connections for gas leaks using a liquid leak ...

Summary

Introduction to the SS2100 r H2 S Analyzer

General Warnings and Cautions

Alerts users to potential hazards and important safety information.

Instructional Symbols

Explains the meaning of symbols used for warnings and instructions within the manual.

How to Use This Manual

Guides users on navigating and utilizing the manual's features and content.

About the Gas Analyzers

Introduces the SS2100r analyzers and their fundamental operating principles.

How the Analyzers Work

Details the tunable diode laser absorption spectroscopy (TDLAS) principle used by the analyzers.

Wavelength Modulation Spectroscopy (WMS) Signal Detection

Describes the WMS technique used for sophisticated signal detection.

Getting Familiar with the Analyzer

Guides users through the analyzer's front, rear panels, and internal compartments.

Analyzer Electronics Assembly

Details the components and layout of the analyzer's main electronics.

Safety Symbols and Labels

Explains safety symbols and labels found on the equipment and in the manual.

Installation Procedures

Contents of the Shipping Box

Lists the expected contents of the analyzer's shipping crate for verification.

Inspecting the Analyzer

Provides instructions for unpacking and inspecting the analyzer for physical damage.

Mounting and Connecting Power

Covers rack mounting procedures and electrical power connection for the analyzer.

Connecting Signals and Gas Lines

Guides on connecting signal, alarm outputs, and sample gas lines to the analyzer.

Calibrating the Analog Output

Guides on calibrating the analyzer's 4-20 mA analog output for accuracy.

Conditioning the Analyzer System

Describes the process of conditioning the analyzer system for stable and accurate readings.

H2 S Scrubber Maintenance

Scrubber Life and Indicators

Discusses factors affecting scrubber lifespan and the function of efficiency indicators.

Servicing the Scrubber

Provides step-by-step instructions for replacing the H2S scrubber and indicator.

Disposal of Used Scrubbers

Advises on the proper disposal procedures for depleted H2S scrubbers.

Specifications

Performance Specifications

Lists key performance metrics including range, repeatability, and response time.

Application and Environmental Data

Details environmental operating ranges and application-specific data.

Electrical and Communication Specifications

Covers electrical input requirements and communication interface details.

Spare Parts Listing

Lists replaceable electronic, conditioning system, and general spare parts.

Troubleshooting and Maintenance

Personnel Safety Risks

Addresses potential hazards and necessary safety actions for personnel during analyzer service.

Common Problems and Causes

Discusses issues like gas leaks, contamination, and environmental factors affecting measurements.

Cleaning Analyzer Components

Provides instructions and necessary tools for cleaning analyzer mirrors.

Instrument Problem Solutions

Lists common instrument problems and their corresponding troubleshooting responses.

Service and Warranty Information

Provides contact details, repair order process, and warranty terms for support.

Endress+Hauser SpectraSensors SS2100r Specifications

General IconGeneral
TypeGas Analyzer
TechnologyTunable Diode Laser Absorption Spectroscopy (TDLAS)
Operating Temperature-20 to +50°C (-4 to +122°F)
EnclosureStainless steel
Measured gasH2O
Measurement RangeApplication dependent
Accuracy±1% of reading or better
Repeatability+/- 1% of reading
Operating PressureUp to 20 bar (290 psi) depending on application
Power Supply24 VDC
Outputs4-20 mA
Communication ProtocolsModbus
MountingWall mount
Enclosure RatingIP66
CertificationsATEX, IECEx