Troubleshooting & Maintenance
Hardware Manual B
5
Tools and Supplies:
Lens cleaning cloth (Cole Parmer
®
EW-33677-00 TEXWIPE
®
Alphawipe
®
Low-Particulate Clean Room Wipes or equivalent)
Reagent-grade isopropyl alcohol (ColeParmer
®
EW-88361-80 or
equivalent)
Small drop dispenser bottle (Nalgene
®
2414 FEP Drop Dispenser
Bottle or equivalent)
Acetone-impenetrable gloves (North NOR CE412W Nitrile Chemsoft™
CE Cleanroom Gloves or equivalent)
Hemostat (Fisherbrand™ 13-812-24 Rochester-Pean Serrated
Forceps)
Bulb blower or dry compressed air/nitrogen
Torque wrench
Permanent ink marker
Non-outgassing grease
Flashlight
To clean the mirrors:
1. Power down the analyzer following the procedure outlined in “To
power down the analyzer” in the Firmware Manual.
2. Isolate the sample conditioning system
(SCS) from the process
sample tap.
The sample cell assembly contains a low-power, 20 mW MAX, CW
Class 3b invisible laser with a wavelength between 800-3000 nm.
Never open the sample cell flanges or the optical assembly unless
the power is turned off.
The sample cell assembly contains a low-power, 20 mW MAX, CW
Class 3b invisible laser with a wavelength between 800-3000 nm.
Never open the sample cell flanges or the optical assembly unless
the power is turned off.
All valves, regulators, switches, etc. should be operated in
accordance with site lock-out/tag-out procedures.