SS2100r H
2
S Analyzer
2–6 4900002189 rev. D 3-7-18
4. For 4-20 mA sink, carefully replace the jumper to connect the center
pin with point “P.”
5. Reconnect power to the analyzer. Confirm the 4 mA (min.) and
20 mA (max.) points (refer to "Calibrating the analog output"
below).
6. Replace and the left side enclosure cover.
Calibrating the analog output
1. Connect a calibrator and digital multi-meter into the circuit.
2. On the analyzer keypad, press #2 (Mode 2), the password (3142)
and *.
3. Continue pressing the * key until the 4-20 mA Test parameter
displays.
Needle nose pliers may be required to remove the jumper.
Figure 2–2 Analyzer 4-20 mA board
JMP1
POTENTIOMETERS
<SET PARAMETER MODE>
AO 4-20 mA Test
0.00000
Enter a value (%)