Hardware Manual A–1
Appendix A: Specifications
Table A–1 General analyzer specifications
1
1. For specific performance specifications, refer to the associated calibration certificate.
Performance
Measurement Range
2
2. Refer to “Customer Service” on page B-11 for alternative ranges.
Refer to the Calibration Certificate
Repeatability Refer to the Calibration Certificate
Response Time
3
3. Software adjustable.
Display updates every 16 seconds
Periodic Scrubber Cycle Duration 90 seconds
Application Data
Environmental Temperature Range 20 C to 30 C (68F to 86F)
Environmental Relative Humidity 5% to 95%, Non-condensing
Altitude Up to 4000 m
Cell Operating Pressure Refer to the Calibration Certificate
Measurement Cell Flow Rate Refer to the Calibration Certificate
Contaminant Sensitivity
None for gas phase glycol, methanol, amines, or
mercaptans.
Electrical & Communications
Input Voltage
4
4. Supply voltage not to exceed 10% of nominal. Transient over-voltages according to Over Voltage
Category II.
120 or 240 VAC 10%, 50/60 Hz; 160W max.
Current (analyzer) 2.5 A max @ 120 VAC
Serial Communication RS-232C (all parameters)
Digital Signals Outputs: (1) General Fault, (1) Hi/Lo Alarm
Analog Output
(1) 4-20 mA Isolated Output, 1200 ohms @ 24
VDC max.
Protocol Modbus Gould RTU, Daniel RTU or ASCII
LCD Display Concentration, cell pressure, cell temperature
Physical Specifications
Size
356 mm H 480 mm W 671 mm D
(14” H 18.9” W 26.4” D)
Weight Approx. 39 kg (86 lbs)
Sample Cell Construction 316L series polished stainless steel - Standard
Area Classification
Certification General Purpose