EasyManua.ls Logo

Endress+Hauser SpectraSensors SS500 - To Connect the Sample Return

Endress+Hauser SpectraSensors SS500
90 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Installation
Hardware Installation and Maintenance Manual 3
13
6. Run stainless steel tubing from the field pressure reducing station
(set for the specified inlet pressure) to the sample supply port of the
SCS. Bend tubing using industrial grade benders, check tubing fit to
ensure proper seating between the tubing and fittings. Fully ream all
tubing ends. Blow out the lines for 10
15 seconds with clean, dry
nitrogen or air prior to making the connection.
7. Connect the inlet tube to the SCS using the 1/4” stainless steel
compression-type fitting provided.
8. Tighten all new fittings 1-1/4 turns with a wrench from finger tight.
For connections with previously swaged ferrules, thread the nut to
the previously pulled up position, then tighten slightly with a wrench.
Secure tubing to appropriate structural supports as required.
9. Check all connections for gas leaks. SpectraSensors recommends
using a liquid leak detector.
Toconnectthesamplereturn
1. Confirm that the atmospheric vent header shut-off valve is closed, if
applicable.
2. Determine appropriate tubing route from the SCS to the
atmospheric vent header.
3. Run stainless steel tubing from the sample return port to the
atmospheric vent header connection.
a. Bend tubing using industrial grade benders, check tubing fit to
ensure proper seating between the tubing and fittings.
b. Fully ream all tubing ends.
c. Blow out the lines for 10–15 seconds with clean, dry nitrogen or
air prior to making the connection.
4. Connect the sample return tube to the SCS using the 1/4” stainless
steel compression-type fitting provided.
5. Tighten all new fittings 1-1/4 turns with a wrench from finger tight.
For connections with previously swaged ferrules, thread the nut to
Do not exceed 10 PSIG (0.7 barg) in sample cell. Damage to cell
may result.
All valves, regulators, switches, etc. should be operated in
accordance with site lock-out/tag-out procedures.

Table of Contents

Related product manuals