Enraf-Nonius B.V. Curapuls 670 Service Manual 1403773
Introduction June 19971-6
Colour specification : EN Custom colours
1)
I: The equipment has a safety earth (ground) connection, and must be connected to
a wall socket with protective earth connection.
2)
BF: The equipment has a floating patient circuit. The Curapuls 670 meets the
requirements of IEC 601 series safety standards.
3)
Copies of certificates or certificates of conformity (COC) are available on request.
1.3.2 European union:
Marking : CE
4)
4)
This equipment complies with all requirements of the Medical Device
Directive (93/42/EEC) and with EMC Guideline 89/336/EEC.
1.3.3 Shortwave therapy
Shortwave frequencies : 27.12 MHz ± 0,6 %
Channels : 2
Output pulse duration : 65, 82, 110, 150, 200, 300, 400
s (±5%)
Pulse frequency : 26, 35, 46, 62, 82, 110, 150, 200, 300, 400, 600 and
800 Hz (± 5%) (600 and 800 Hz Circuplode E only,
single use)
Pulse power : 0 - 200 W (±20%) in steps of 5W into 50
6
Mean power : 0 - 64 W (48 and 64 W Circuplode E only) into 50 6
Contact control threshold level : 80 % of initial value
Timer : 0 - 30 minutes (±5%)