EasyManuals Logo

Enraf Nonius Radarmed 950+ User Manual

Enraf Nonius Radarmed 950+
Go to English
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #59 background imageLoading...
Page #59 background image
ENGLISH
EN109-1435752-42 Page 27 of 29
Guidance and manufacturer’s declaration – electromagnetic immunity
The RADARMED is intended for use in the electromagnetic environment specified below. The
customer or the user of the RADARMED should assure that it is used in such an environment.
IMMUNITY
test
IEC 60601 test level
Compliance level
Electromagnetic
environment guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-
2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least 30
%.
Electrical fast
transient/burst
IEC 61000-4-
4
± 2 kV for power supply
lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
Surge
IEC 61000-4-
5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-
11
<5 % UT
(>95 % dip in UT) for
0,5 cycle
40 % UT
(60 % dip in UT) for 5
cycles
70 % UT
(30 % dip in UT) for 25
cycles
<5 % UT
(>95 % dip in UT) for 5
s
<5 % UT
(>95 % dip in UT) for
0,5 cycle
40 % UT
(60 % dip in UT) for 5
cycles
70 % UT
(30 % dip in UT) for 25
cycles
<5 % UT
(>95 % dip in UT) for 5
s
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the RADARMED
requires continued operation
during power mains
interruptions, it is
recommended that the
RADARMED be powered from
an uninterruptible power supply
or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-
8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.
NOTE UT is the AC mains voltage prior to application of the test level.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Enraf Nonius Radarmed 950+ and is the answer not in the manual?

Enraf Nonius Radarmed 950+ Specifications

General IconGeneral
BrandEnraf Nonius
ModelRadarmed 950+
CategoryMedical Equipment
LanguageEnglish

Related product manuals