12 Maintenance and troubleshooting
Cleaning and Disinfection
Before performing any user maintenance, switch off the device and disconnect the plug from
the mains supply.
Do not spray the cleaning agent directly on the glass panel.
Do not use cleaning agents that contain strong alkalis, lye, acid, detergents with fluoride or
detergents with ammonia.
Use no liquid detergents, these can damage the device.
Do not allow any liquids to penetrate the device or its accessories while cleaning and
disinfecting!
Dry all sockets and connectors that have become wet before any further use!
Mechanical shocks /or vibrations must be avoided while cleaning ultrasound head!
Do not drop the ultrasound applicator!
Device
Display Panel
Ultrasound
Applicator
Troubleshooting
When the apparatus is turned on, it will first execute a self test. When an error is detected, both during
the self test and during normal operation, a pop-up screen will appear on the display. When the error
is displayed, all outputs will be disabled. When this situation occurs, remove all cables and switch the
apparatus off and on again. When the error re-appears, stop using the device and contact your
supplier.
Technical Maintenance
Electrical safety of the device relies on a proper earthed electrical connection via the power
cord. It is therefore necessary to have this connection checked annually.
To ensure continued compliance with the 21 CFR 1050.10 standard, this unit should be
adjusted and safety tested once each year. Procedures laid down in the service manual
should be followed. This may be carried out by your supplier, or by another agency,
authorized by the
manufacturer. It is also recommended that a service history record is maintained. In some
countries this is even obligatory.
Use of controls or adjustments or performance of procedures other than those specified in
the service manual may result in hazardous exposure to ultrasonic energy.

Other manuals for Enraf Nonius Sonopuls 190

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Enraf Nonius Sonopuls 190 and is the answer not in the manual?

Enraf Nonius Sonopuls 190 Specifications

General IconGeneral
BrandEnraf Nonius
ModelSonopuls 190
CategoryMedical Equipment
LanguageEnglish

Related product manuals