TROUBLESHOOTING
Fluid Warmer Operator’s Manual MN-39935 Rev 1 3/19
33
What to do if the Over Temperature Indicator
Light Illuminates
Background When a temperature of 10°F (6°C) greater than the set-point temperature occurs,
the controller sounds an alarm. The over temp status indicator LED
flashes
.
The over temp status indicator LED will turn off when the actual temperature
returns to the set-point temperature.
Procedure
If the over temperature indicator light illuminates, do the following.
Step Action
1. Press the ON/OFF button to acknowledge the over temperature.
The controller will reset.
2. Make sure the fluids are the correct temperature before using.
Result Resume operation of the warmer.