7.4 Required aids
7.5 Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
2. Move the movable parts back and forth (at least 5 times).
3. Mount the irrigation tube on the shaft tube.
4. Brush the inside of the shaft tube with a round brush (at
least 5 times).
5. Completely fill a disposable syringe with fresh water and
rinse the lumina (at least 5 times).
Reprocessing step Items Needed
Precleaning Soft plastic brush/soft cloth
Round brush (such as Art. No. 20191-
279)
Irrigation tube (Art. No. 20195-201)
50 ml disposable syringe
Manual cleaning/
disinfection
Soft plastic brush/soft disposable cloth
(low particle)
Round brush (such as Art. No. 20191-
279)
Irrigation tube (Art. No. 20195-201)
50 ml disposable syringes (4 pc)
Compressed air gun (2 bar maximum)
Machine cleaning/
disinfection
Irrigation tube (Art. No. 20195-201)
OBJ_DOKU-163748-014.fm Seite 28 Donnerstag, 19. September 2019 3:10 15

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Erbe BiClamp 110/340 and is the answer not in the manual?

Erbe BiClamp 110/340 Specifications

General IconGeneral
BrandErbe
ModelBiClamp 110/340
CategoryMedical Equipment
LanguageEnglish

Related product manuals