ADVANCED OPERATION - Special functions: Global parameters
Table of contents
Parameter Function Range Default
LOSS EST. LEV Gives you the option of deciding 0 to 2 2
in what cases you want the splicer
to estimate the splice loss.
0 = No loss estimation
1 = Estimation only when in Automode
2 = Always do loss estimation
SHOW LOSS TO When the estimated loss exceeds 0.0 to 49.9 dB 2.0 dB
this value the splicer will display
BAD SPLICE.
MACROBENDING A proportionalizing factor that is 0.0 to 9.9 0.3
used to optimize estimation of
loss due to macrobending.
AUTOXCT OFFSET Gives the offset to the set splicing -90 to +90 pixels 0 pixels
position (see also AUTO ARC
CENTER on p. 56).
GLOBAL XCT Allows you to change the splicing 150 to 360 pixels 255 pixels
position (see also SET CENTER
POSITION on p.55)
WEIGHT FACTOR A factor that sets the grade of 0.00 to 0.99 0.25
impact for measured data in the (0.00 = low
equation used by the Automatic 0.99 = high)
Arc Re-centering function.
RTC CURR LOL Low limit for RTC current. 0 to 29.9 mA 14.5 mA
RTC CURR HIL High limit for RTC current. 0 to 29.9 mA 15.5 mA
SPLICES TOTAL: The total number of splices made 0 to xx
by the splicer: UNEDITABLE.
SPLICES: Splice counter. 0 to xx
RESET SPLICES? Resets splice counter to 0. Use YES/NO NO (=0)
after you have replaced electrodes.
WAIT TIME Time of inaction before splicer 0.0 to 4.9 MN 1.0 MN
turns off automatically.
ELECTROD WARNING Number of splices before splicer 00 to 99 HS 20 HS
warns you that it is time to (HS = hundred (=2000 splices)
replace the electrodes. splices)
TURN OFF Enable/Disable the function that YES/NO YES (=1)
automatically turns off the splicer
YEAR The current year. 00 to 99 set in factory
MONTH The current month. 00 to 12 set in factory
DAY The current day. 00 to 31 set in factory
HOUR The current hour. 00 to 24 set in factory
MIN The current minute. 00 to 59 set in factory
Estimation parameters
Current
CPU settings Offset parameters